Determination of oligonucleotides by XtimateC18 HPLC Column

Test Report

Chromatography conditions:

Column 

Welch Xtimate® C18(4.6×250mm,5um)

Mobile Phase

A: Phosphate buffer/methanol=95/5; B: Phosphate buffer/methanol=75/25

Column Temperature

25 ℃

Detector

UV

Injection Volume

10μL

Flow Rate

1.0 ml/min

Detect wavelength

260nm

Elution procedure

Time (min)

A(%)

B(%)

0

100

0

20

100

0

30

50

50

35

0

100

40

0

100

40.1

100

0

55

100

0


Note

/

Chromatogram and Data:

Retention time 

K

Area(nAU*S)

Height(mAU)

Symmetry Factor

Peak width(min)

Number of theoretical plates

Resolution

Selectivity

5.925

-

19.6584

2.97282

1.04

0.1033

18216

-

-

7.441

-

65.75619

6.43044

1.05

0.1600

11985

6.76

1.26

11.945

-

1086.99316

73.28335

1.03

0.2275

15279

13.66

1.61

18.117

-

2350.23511

116.62537

1.12

0.3109

18803

13.47

1.52

Conclusion:

Under these HPLC conditions. Welch Xtimate® C18(4.6×250mm,5um)can meet the customer's requirements.