Test Report

Chromatography conditions:

Column 

Welch Ultisil®XB-CN4.6×250mm5μm

Welch Topsil® CN4.6×250mm5μm

Mobile Phase

n-heptane/isopropyl alcohol/methylene chloride/acetonitrile=80/10/6/4

Column Temperature

40℃

Detector

UV

Injection Volume

5μL

Flow Rate

1.0 ml/min

Detect wavelength

240 nm

Note

/

Chromatogram and Data: 

1) Welch Ultisil® XB-CN 202360607 test solution detection spectrum

 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

11.072

2749.687

174.883

0.0

1.159

12823

2) Welch Ultisil® XB-CN 202303118 test solution detection spectrum

Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

8.958

9.458

0.773

5.8

1.594

11605

2.

N.a

11.098

5322.349

321.348

0.0

1.165

11617

3

Sum

 

 

 

 

 

 

  • 3) Welch Ultisil® XB-CN spiked test solution detection chart


 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

8.978

39.772

3.278

6.0

1.041

12764

2.

N.a

11.070

3024.908

196.887

0.0

1.143

13247

3

Sum

 

 

 

 

 

 

  • 4) Topsil® CN 202360607 Test sample solution detection chart

Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

10.397

3028.524

200.291

0.0

1.232

12193

2

Sum

 

 

 

 

 

 

  • 5) Topsil® CN 202303118 spiked test solution detection chart

Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

8.040

10.595

1.118

7.2

1.114

16762

2

N.a.

10.338

5283.361

336.957

0.0

1.238

11259

 

Sum

 

 

 

 

 

 

  • 6) Topsil® CN spiked test solution detection chart


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

Peak height(mAu)

Resolution(EP)

Tailing factor

Number of plates(EP)

1

N.a.

8.065

42.304

3.880

7.0

1.235

13307

2

N.a.

10.353

2984.087

196.951

0.0

1.236

12014

 

Sum

 

 

 

 

 

 

Conclusion:

Under these chromatographic conditionsUltisil ® XB-CN4.6×250mm5μm)and Topsil® CN4.6×250mm5μm), can meet the customers requirements.