Welcome to our new website!

Determination of mannitol content with Ultisil HILIC-Amide

Chromatography conditions:

Column

Ultisil®HILIC Amphion II(4.6×250mm,5um)  

Mobile Phase

Acetonitrile/water=75/25

Injection Volume

10 μL

Flow Rate

1.0 ml /min

Detector

ELSD

Column Temperature

30℃

Detection wavelength

ELSD detector temperature: 80℃, carrier gas flow rate: 1.8L/min

Note

/

 

Chromatogram and Data: 

  1. Blank solution detection chart

2. Mannitol single standard solution detection chart

Retention time(min) 

K

Area(mV*s)

Height(mV)

Symmetry Factor

Width (min)

Number of plates(USP)

Resolution(USP)

Selectivity

11.355

-

2691.67212

216.02324

1.11

0.1881

20184

-

--

3. Detection pattern of sample solution 1 (used for method development)

Retention time(min) 

K

Area(mV*s)

Height(mV)

Symmetry Factor

Width (min)

Number of plates(USP)

Resolution(USP)

Selectivity

11.162

-

6483.72754

440.26373

1.03

0.2277

13311

-

--

4. Sample solution 2 (for testing) detection pattern

Retention time(min) 

K

Area(mV*s)

Height(mV)

Symmetry Factor

Width (min)

Number of plates(USP)

Resolution(USP)

Selectivity

11.122

-

5244.35352

356.00610

1.14

0.2275

13241

-

--

Conclusion:
Using Welch Ultisil® HILIC-Amide (4.6×250mm, 5μm) chromatography column under these chromatographic conditions, the peak shape and retention of mannitol are excellent, with no interference from other substances, meeting the customer's detection requirements.