Welcome to our new website!

Determination of the content of formic acid, acetic acid, methanol, ethanol, and n-propanol in alkali solution by Xtimate Sugar-H

Test Report

Chromatography conditions:

Column

Xtimate® Sugar-H(7.8×300mm,5μm)

Mobile Phase

0.025mol/L sulfuric acid

Detector

RID

Injection Volume

20 μL

Flow Rate

0.6 ml /min

Column Temperature

60℃

Detector Temperature

35℃

Note

/

Chromatogram and Data: 

  1. Detection chart of test solution

Retention time(min) 

K

Area(nRIU*s)

Height(nRIU)

Symmetry Factor

Width (min)

Number of plates

Resolution

Selectivity

0.076

-

19.63885

3.60375

0.92

0.0900

4

-

--

13.220

-

5085.25146

366.33725

0.88

0.2088

22205

51.69

174.86

14.217

-

8615.93848

608.80865

0.91

0.2160

24011

2.76

1.08

18.617

-

2.75426e4

1542.47729

0.86

0.2808

24358

10.41

1.31

21.331

-

9.36549e4

4570.89697

0.94

0.3168

25123

5.34

1.15

26.876

-

9803.13672

376.06458

0.95

0.4032

24618

9.05

1.26

 

Conclusion:
Under these chromatographic conditions, Welch Xtimate® Sugar-H7.8×300mm5μm) can meet the detection requirements.

 