Welcome to our new website!

Determination of unknown sample content using Ultisil HILIC Amphion II

Test Report

Chromatography conditions:

Instrument Info

Shimadzu LC-20AD

Column

Ultisil®HILIC Amphion II(4.6×250mm,5um)  

Mobile Phase

Acetonitrile/5.5g/L ammonium acetate (use acetic acid to adjust pH to 6.0) =65/35

Injection Volume

5 μL

Flow Rate

0.8 ml /min

Detector conditions

Excitation wavelength: 274nm; Emission wavelength: 312nm;

Column Temperature

50℃

Note

/

Chromatogram and Data: 

  1. Detection amplification chart of mixed standard solution

Retention time(min) 

Area

Height

Number of theoretical plates(USP)

Tailing Factor

Resolution(USP)

8.809

8125562

684851

14662

1.234

--

2. Detection chart of test solution


Retention time(min) 

Area

Height

Number of theoretical plates(USP)

Tailing Factor

Resolution(USP)

8.804

9418389

803126

14552

1.210

--

 

Conclusion:
Under these chromatographic conditions, Welch Ultisil®HILIC Amphion II(4.6×250mm,5um)can meet the detection requirements.

Post a comment