Xtimate C8 - Vitamin B12 and its impurities analysis method development report

Xtimate C8 - Vitamin B12 and its impurities analysis method development report

Test Report

Chromatography conditions:

Column 

Xtimate C83.0×250mm3μm

Mobile Phase

0.028mol/L disodium hydrogen phosphate solution (pH=3.1): methanol = 82:18

Column Temperature

40℃

Detector

UV

Injection Volume

20μL

Flow Rate

0.6 ml/min

Detect wavelength

361 nm

Elution procedure

Time (min)

0.028mol/L disodium hydrogen phosphate solution (pH=3.1): methanol = 82:18

0

100%

45

100%


Note

/

Chromatogram and Data: 

  • 1) Analysis chart of vitamin B12 impurity spiked test solution:


 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

Impurity H

3.738

0.885

4.7

0.20

24.19

10112

1.33

2

Vitamin B12

12.353

413.471

1052.9

92.03

11.53

7632

1.60

3

Impurity A

19.318

6.741

87.7

1.50

2.47

14318

0.90

4

Impurity C

21.038

15.247

171.1

3.39

4.86

12556

0.93

5

Impurity D

24.795

5.473

218.4

1.22

1.50

15403

0.88

6

Impurity E

25.968

0.447

19.6

0.10

4.11

17983

0.97

7

Impurity B

29.480

7.003

144.0

1.56

N.a.

15814

1.06

Sum


 

449.267

 

100.0

48.65

 

 


  • 2) Analysis result reproducibility verification superposition spectrum:


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

Impurity H

3.738

0.885

4.7

0.20

24.19

10112

1.33

2

Vitamin B12

12.353

413.471

1052.9

92.03

11.53

7632

1.60

3

Impurity A

19.318

6.741

87.7

1.50

2.47

14318

0.90

4

Impurity C

21.038

15.247

171.1

3.39

4.86

12556

0.93

5

Impurity D

24.795

5.473

218.4

1.22

1.50

15403

0.88

6

Impurity E

25.968

0.447

19.6

0.10

4.11

17983

0.97

7

Impurity B

29.480

7.003

144.0

1.56

N.a.

15814

1.06

Sum


 

449.267

 

100.0

48.65

 

 

  • 3) Multi-needle analysis of pressure stability overlay:



Conclusion:

Vitamin B12 and its impurities were analyzed under these chromatographic conditions using a Welch column Xtimate C8 (3.0×250mm, 3μm). Vitamin B12 and impurities A, C, D, E and B eluted in sequence with a separation degree of no less than 1.5. After multiple injections of the spiked solution of the test sample, the results showed good reproducibility and stable pressure, which met the analysis requirements.