What is Ghost Peak?
The peak appears erratically like ghost in chromatographic separation, especially during gradient elution or long-period operation.
Where is Ghost Peak from?
- Water, with impurities Purification System, polluted or poorly functioning
- Storage Containers, polluted or breeding bacteria
- Mobile Phase Additives, like salts acids and alkalis
- Instrument, polluted after long-period use
- Other organic pollutants
Welch Ghost-Buster Column can effectively remove impurities with low polarity and thus prevents the interference from all kinds of ghost peaks. It is installed between gradient mixer and injector, which helps remove not only the impurities in mobile phase, but impurities in mixer and pipelines as well.
Operation Principles
Unlike in-line filters which removes only solid particles but not organic pollutants, Welch Ghost-Buster column provides strong adsorption to weak-polar and non-polar organic impurities, without changing the composition of mobile phase thus to purify both mobile phase and system, remove most ghost peaks and extend lifetime of column and system.
Application
Column: Ultisil XB- C18, 4.6x250 mm, 5 um
Flow Rate: 1.0 mL/min
Injection Volume: 10 uL
Detection: 210 nm
Temperature: 40℃
Sample Preparation: Ultra-pure water
Mobile Phase A: Ultra-pure water
Mobile Phase B: Acetonitrile
Precautions
- Install the column between Mixer and Injector. Being installed after injector would cause strong adsorption to samples and affect analysis.
- For new analytical columns, flush Ghost-Buster column with 80% methanol solution at 1ml/min for 20min before new column switching to the system.
- Not all impurities can be adsorbed by Ghost-Buster column.
- Ion-pair solvents in mobile phase, would be adsorbed by Ghost-Buster column and affect retention and peak shape. Please use with caution under such mobile phases.
- Column lifetime depends on analytical conditions, mobile phase and solvent purity. Routine change of Ghost-Buster column is suggested to ensure performance.
- Ghost-Buster column is rather a purification part to the system, to filtrate impurities and protect column and system
- Before and after using buffer salt mobile phase, flush column with high-ratio water to transit, thus to avoid buffer salting out and blocking the column.
- When Ghost-Buster column shows unsatisfying performance, try disconnect the outlet of the column and flush with100% acetonitrile.
Q&A
Q: For different samples and gradient conditions, should the Ghost-Buster column be removed or changed?
ID: 2.1mm
Length: 33mm
Max.Pressure: 40Mpa
What is Ghost Peak?
The peak appears erratically like ghost in chromatographic separation, especially during gradient elution or long-period operation.
Where is Ghost Peak from?
- Water, with impurities Purification System, polluted or poorly functioning
- Storage Containers, polluted or breeding bacteria
- Mobile Phase Additives, like salts acids and alkalis
- Instrument, polluted after long-period use
- Other organic pollutants
Welch Ghost-Buster Column can effectively remove impurities with low polarity and thus prevents the interference from all kinds of ghost peaks. It is installed between gradient mixer and injector, which helps remove not only the impurities in mobile phase, but impurities in mixer and pipelines as well.
Operation Principles
Unlike in-line filters which removes only solid particles but not organic pollutants, Welch Ghost-Buster column provides strong adsorption to weak-polar and non-polar organic impurities, without changing the composition of mobile phase thus to purify both mobile phase and system, remove most ghost peaks and extend lifetime of column and system.
Application
Column: Ultisil XB- C18, 4.6x250 mm, 5 um
Flow Rate: 1.0 mL/min
Injection Volume: 10 uL
Detection: 210 nm
Temperature: 40℃
Sample Preparation: Ultra-pure water
Mobile Phase A: Ultra-pure water
Mobile Phase B: Acetonitrile
Precautions
- Install the column between Mixer and Injector. Being installed after injector would cause strong adsorption to samples and affect analysis.
- For new analytical columns, flush Ghost-Buster column with 80% methanol solution at 1ml/min for 20min before new column switching to the system.
- Not all impurities can be adsorbed by Ghost-Buster column.
- Ion-pair solvents in mobile phase, would be adsorbed by Ghost-Buster column and affect retention and peak shape. Please use with caution under such mobile phases.
- Column lifetime depends on analytical conditions, mobile phase and solvent purity. Routine change of Ghost-Buster column is suggested to ensure performance.
- Ghost-Buster column is rather a purification part to the system, to filtrate impurities and protect column and system
- Before and after using buffer salt mobile phase, flush column with high-ratio water to transit, thus to avoid buffer salting out and blocking the column.
- When Ghost-Buster column shows unsatisfying performance, try disconnect the outlet of the column and flush with100% acetonitrile.
Q&A
Q: For different samples and gradient conditions, should the Ghost-Buster column be removed or changed?
ID: 2.1mm
Length: 33mm
Max.Pressure: 40Mpa