Anthocyanin on Ultisil PFP(b)
Mixed standard solution
Gradient Program
Time(min) | %A | %B |
0 | 92 | 8 |
5 | 90 | 10 |
22 | 90 | 10 |
30 | 85 | 15 |
35 | 80 | 20 |
40 | 65 | 35 |
45 | 50 | 50 |
50 | 35 | 65 |
55 | 30 | 70 |
55.1 | 92 | 8 |
65 | 92 | 8 |
Column | Welch Ultisil PFP 4.6×250mm,5μm (P/N: 00224-31043) |
Mobile Phase | Mobile phase A: 5.0% formic acid in water Mobile phase B: methanol |
Detector | 530nm |
Temperature | 30℃ |
Flow Rate | 0.8 ml/min |
Injection Volume | 10µl |