Azithromycin on Ultisil AQ-C18
Related Substance System Suitability
Column | Ultisil® AQ-C18 4.6×250 mm, 5 μm P/N: 00207-31043 |
Detection | 210 nm |
Temperature | 30 ℃ |
Flow Rate | 1.1 ml/min |
Injection Volume | 50 µl |
Mobile Phase
A: 0.05mol/L dipotassium phosphate solution (Adjust pH to 8.2 with 20% phosphoric acid solution): Acetonitrile=45:55 B: Methanol
Time | A% | B% |
0 | 75 | 25 |
35 | 95 | 5 |
64 | 95 | 5 |
65 | 75 | 25 |
71 | 75 | 25 |