Carbidopa TabletContent Test Report on Ultisil LP-C18

System adaptation test

No. Rt (min) Area (uAU*s) Height (uAU) Area (%) Plates Resolution Tailing Factor
1 23.241 2349639 71490 49.76 11478 0.00 1.322
2 25.374 2372414 64898 50.24 11033 2.33 1.361
Total: 4722053 136388 100.00

Specific method:

Column:Ultisil® LP-C18, 5μm, 4.6×250mm

Detector:280nm

Mobile phase: Ethanol: 0.05M sodium dihydrogen phosphate adjusted with phosphoric acid to pH 2.7 (5:95)

Flow rate: 1.0ml/min

Injection Volume: 20ul