Determination of various organic acids using Ultisil OAA (1)

Determination of various organic acids using Ultisil OAA (1)

Test Report

Sample Information

Sample name 

16 organic acids

No.

W20190516-001

Sample weight

/

 

/

Sample collection date


201905/16

Test period

2019/05/16-05/24

Sample description

/

Test requirements

Test ingredients

16 organic acids

Reference standards

Reference standards

/

Instrument information

Test instrument

High performance liquid chromatograph

Instrument model

Thermo Fisher U3000


Chromatography conditions:

Column 1

Welch Ultisil® OAA4.6×300mm

Mobile Phase

Time (min)

0.01mol/L potassium dihydrogen phosphate solution (%)

Methanol (%)

0

100

0

15

100

0

30

95

5

35

80

20

45

60

40

45.1

100

0

65

100

0

 

Column Temperature

32 ℃

Detect wavelength

210nm

Injection Volume

μL

Flow Rate

0.5ml/min

Note

/

Preparation of mobile phase:

0.01mol/L potassium dihydrogen phosphate solution: weigh 1.36g potassium dihydrogen phosphate, dissolve in 1000ml water, adjust pH to 2.5 with phosphoric acid,

filter, and obtain.

Methanol: take chromatographic pure methanol, and obtain.

Preparation of sample solution:

Solvent: water;

Solution preparation: take each reference substance, prepare a mixed solution with water, mix well, and obtain

No.

Name

 

Figure

Retention time

Concentration

(mg/ml)

No.

Name

 

Figure

Retention time

Concentration

(ug/ml)

 1 

Oxalic acid

7.675

0.20

9

Acetic acid

15.035

2.0

2

Tartaric acid

8.707

1.50

10

Maleic acid

20.760

0.15

3

Tartaric acid

9.073

1.50

11

Citric acid

22.148

5.0

4

Quinic acid

9.340

0.80

12

Fumaric acid

25.472

0.06

5

Formic acid

10.927

0.24

13

Citraconic acid

29.033

0.2

6

Pyruvic acid

Malic acid

11.460

3.50

14

Acrylic acid

31.653

0.2

7

Ascorbic acid

13.120

0.50

15

Propionic acid

35.857

8.0

8

Lactic acid

14.095

0.50

16

Citraconic acid

39.895

0.15

 

Chromatogram and Data: 

  • 1. Mixed control solution



Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

peak height(mAU)

Number of plates(EP)

Asymmetry(EP)

Resolution(EP)

1

Oxalic acid

7.675

40.851

378.465

33761

1.21

5.89

2

Tartaric acid

8.707

37.535

316.810

35807

0.98

1.98

3

Tartaric acid

9.073

8.475

73.218

37806

0.99

1.32

4

Quinic acid

9.340

9.953

70.697

29417

1.37

4.23

5

Formic acid

10.927

24.224

73.867

6676

0.95

1.36

6

Pyruvic acid

Malic acid

11.460

38.520

245.564

33615

1.18

6.19

7

Ascorbic acid

13.120

73.721

395.865

33226

0.98

3.29

8

Lactic acid

14.0895

3.561

18.295

34052

1.02

2.67

9

Acetic acid

15.035

20.785

80.654

22560

1.39

11.93

10

Maleic acid

20.760

243.295

691.126

21848

1.94

2.35

11

Citric acid

22.148

73.297

189.673

20254

1.99

5.52

12

Fumaric acid

25.472

137.295

368.908

30263

0.95

5.43

13

Citraconic acid

29.033

197.471

432.214

25336

2.00

3.67

14

Acrylic acid

31.653

125.177

279.656

32690

1.06

5.00

15

Propionic acid

35.857

56.502

90.696

21147

1.89

4.90

16

Citraconic acid

39.865

139.454

327.895

57221

1.24

N.a.

Sum

 

 

1230.116

4033.603

475679.00

21.18

 

Conclusion:

Under this chromatographic conditionWelch Ultisil® OAA (4.6×300mm)can meet the detection requirements