Fasudil Hydrochloride on Ultisil XB-C18 and Xtimate C18
Sensitivity solution
# | RT | Compound | Area | Height | Area% | Plates(USP) | Resolution(USP) | S/N |
1 | 4.698 | Fasudil hydrochloride | 4305 | 684 | 100.00 | 11116 | -- | 71.04 |
Total: | 4305 | 684 | 100.00 |
Column | Welch Ultisil XB-C18 4.6×250mm,5μm (P/N: 00201-31043) Welch Xtimate C18 4.6×250mm,5μm (P/N: 00101-21043) |
Mobile Phase | Triethylamine buffer: methanol = 50:50 |
Detector | 275nm |
Temperature | 40℃ |
Flow Rate | 1ml/min |
Injection Volume | 20ul |
Notes | The column temperature of this experiment is 40 ° C, which can make the column pressure smaller. |