Ultisil Cellu-DR-MBLSW Isomer Analysis Method Development Report

Ultisil Cellu-DR-MBLSW Isomer Analysis Method Development Report

Test Report


Chromatography conditions:

Column 

Ultisil Cellu-DR4.6×300mm,5μm

Mobile Phase 

Phase A: 0.1% formic acid in water Phase B: 0.1% formic acid in acetonitrile

Injection Volume

10 μL

Column temperature

25 ℃

Flow Rate

0.8 ml/min

Detector

UV

Detect wavelength

254 nm

Elution procedure

Time (min)

Mobile Phase A%

Mobile Phase B%

0

65

35

55

65

35


  • 1. Spectra and data
  • 1) Blank solvent analysis spectrum:


  • 2) 0361167 Analysis spectrum of control solution:


 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

0361167

32.148

234.418 

380.0

100.00

N.a.

8243

1.28

Sum


 

234.418 

 

100.00

0.00

 

 



  • 3) 0355439 Control solution analysis spectrum:


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

0355439

45.703

280.178

40.1

100.00

N.a.

4906

0.77

Sum


 

28.178

 

100.00

0.00

 

 


  • 4) 0355424 Control solution analysis spectrum:

 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

0355424

36.302

271.456

146.3

100.00

N.a.

7990

1.45

Sum


 

271.456

 

100.00

0.00

 

 

  • 5) 0355437 Analysis spectrum of control solution:


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution
(EP)

Number of plates(EP)

Asymmetry
(EP)

1


36.558

25.210

62.7

8.66

1.96

8730

N.a.

2

0355437

40.378

265.876

411.7

91.34

N.a.

4791

0.81

Sum


 

291.066

 

100.00

1.96

 

 

 

  • 6) Mixed control solution analysis spectrum:


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution
(EP)

Number of plates(EP)

Asymmetry
(EP)

1

0361167

32.392

53.865

135.0

19.62

2.91

8720

1.18

2

0355424

36.682

71.541

155.7

26.06

2.08

8793

1.32

3

0355437

40.775

77.732

39.4

28.31

2.08

4669

N.a.

4

0355439

46.067

71.407

30.0

26.01

N.a.

4621

N.a.

Sum


 

274.545

 

100.00

7.07

 

 


Conclusion:

Welch column Ultisil Cellu-DR (4.6×300mm, 5μm) was used to analyze the four substances under this chromatographic condition. The separation between them was greater than 2, which met the analysis requirements.