Ultisil XB-C8-Eribulin Mesylate Injection Analytical Method Development Report

Ultisil XB-C8-Eribulin Mesylate Injection Analytical Method Development Report

Test Report

Chromatography conditions:

Column 

Welch Utisil XB-C84.6×150mm,3μm300A

Mobile Phase

Phosphate buffer (pH=7.0)-acetonitrile (73:27)

Column Temperature

30℃

Detector

UV

Injection Volume

50μL

Flow Rate

1 ml/min

Detect wavelength

200 nm

Elution procedure

Time (min)

Phosphate buffer (pH=7.0)-acetonitrile (73:27)

0

100%

45

100%


Note

/

Chromatogram and Data: 

  • 1) Blank solvent analysis spectrum:


 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

Sum


 

0.000

 

0.00

0.00

 

 


  • 2) System suitability solution analysis spectrum full range


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

impA

4.668

6.452

42.1

1.49

12.54

14973

1.24

2

ER

10.557

404.741

496.4

93.77

10.16

2864

4.91

3

impD

17.203

12.065

47.6

2.80

2.77

17320

1.25

4

impE

18.682

6.963

18.4

1.61

12.84

18749

1.14

5

ImpG

27.308

1.408

2.5

0.33

N.a.

18576

1.08

Sum



431.628

 

100.00

38.31

 

 

  • 3) System suitability solution analysis spectrum magnification range:

 Serial number 

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

impA

4.668

6.452

42.1

1.49

12.54

14973

1.24

2

ER

10.557

404.741

496.4

93.77

10.16

2864

4.91

3

impD

17.203

12.065

47.6

2.80

2.77

17320

1.25

4

impE

18.682

6.963

18.4

1.61

12.84

18749

1.14

5

ImpG

27.308

1.408

2.5

0.33

N.a.

18576

1.08

Sum



431.628

 

100.00

38.31

 

 

  • 4) Batch Verification 1303.02 Batch System Suitability Solution Analysis Spectrum:


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

impA

4.665

6.460

40

.9

1.49

12.52

14968

1.22

2

ER

10.543

405.781

482.4

93.75

10.14

2859

4.83

3

impD

17.187

12.243

48.8

2.83

2.77

171780

1.26

4

impE

18.663

7.041

19.3

1.63

13.00

18737

1.12

5

ImpG

27.292

1.324

2.5

0.31

N.a.

19311

1.07

Sum



432.849

 

100.00

38.44

 

 

  • 5) Batch Validation 1303.03 Batch System Suitability Solution Analysis Spectrum


Serial number

Peak name 

Retention time (min)

Peak Area(mAU*min)

S/N

Relative peak area(%)

Resolution(EP)

Number of plates(EP)

Asymmetry(EP)

1

impA

4.668

6.452

42.1

1.49

12.54

14973

1.24

2

ER

10.557

404.741

496.4

93.77

10.16

2864

4.91

3

impD

17.203

12.065

47.6

2.80

2.77

17320

1.25

4

impE

18.682

6.963

18.4

1.61

12.84

18749

1.14

5

ImpG

27.308

1.408

2.5

0.33

N.a.

18576

1.08

Sum



431.628

 

100.00

38.31

 

 


Conclusion:

Welch column Ultisil XB-C8 (4.6×150mm, 3μm, 300A) was used to analyze the system suitability solution under this chromatographic condition. The retention time of the main peak was 10.6min, the separation between the main substance and the impurity was greater than 1.5, and the separation between impurity D and impurity E was greater than 2, which met the analysis requirements.