Test Report
Chromatography conditions:
Column |
Ultisil XB-NH2(4.6×250mm,5μm) |
||||||
Mobile Phase |
10.89g/L Potassium dihydrogen phosphate solution |
||||||
Column Temperature |
35℃ |
||||||
Detection pool temperature |
35℃ |
||||||
Detector |
RID |
||||||
Injection Volume |
20μL |
||||||
Flow Rate |
1.0 ml/min |
||||||
Elution procedure |
|
||||||
Note |
/ |
Chromatogram and Data:
- 1. System suitability solution analysis spectrum full range:
- 2. System suitability solution analysis spectrum part of the range:
- 3. Preparation solution analysis spectrum full range
Conclusion:
Using the Welch Ultisil XB-NH2 column (4.6×250mm, 5μm) under these chromatographic conditions to analyze the system suitability solution, the resolution between the main substance and the four impurities is not less than 1.5, meeting the analytical requirements.