Ultisil XB-SCX-Urea & Bifonazole Analysis Method Development Report

Ultisil XB-SCX-Urea & Bifonazole Analysis Method Development Report

Test Report

Chromatography conditions:

Column 

Welch Ultisil XB-SCX(4.6×250mm,5μm)

Mobile Phase

 Phase A : 6mmol/L ammonium acetate 0.03% TFA aqueous solution Phase B: chromatographic grade acetonitrile

Column Temperature

25℃

Detector

UV

Injection Volume

10μL

 Flow Rate 1

1.0 ml/min

Flow Rate 2

1.3 ml/min

Detect wavelength

195 nm

Elution procedure

Time (min)

Phase A(%)

Phase B(%)

0

20

80

20

20

80


Note

/

Chromatogram and Data: 

1) Blank solvent analysis spectrum



2) Urea reference solution analysis spectrum:



  • 3) Analysis spectrum of bifonazole control solution:


  • 4) Mixed control solution analysis spectrum:

  • 5) Analysis spectrum of mixed control solution at a flow rate of 1.3ml/min:

  • 6) Sensitivity solution analysis spectrum at a flow rate of 1.3ml/min


Conclusion:

Using the Welch Ultisil XB-SCX column (4.6×250mm, 5μm) under these chromatographic conditions to analyze urea and bifonazole, the resolution between them is greater than 1.5. At a flow rate of 1.3 ml/min, the retention times for urea and bifonazole are 4.95 minutes and 11.1 minutes, respectively, meeting the analytical requirements.