Test Report
Chromatography conditions:
Column |
Xtimate C8(3.0*250mm*3μm) |
||||||
Mobile Phase |
0.028mol/L disodium hydrogen phosphate solution (pH=3.1): methanol = 82:18 |
||||||
Column Temperature |
40℃ |
||||||
Detector |
UV |
||||||
Injection Volume |
20μL |
||||||
Flow Rate |
0.6 ml/min |
||||||
Detect wavelength |
361 nm |
||||||
Elution procedure |
|
||||||
Note |
/ |
Chromatogram and Data:
- 1) Test solvent spectrum
2) Reproducibility verification overlay graph:
Conclusion:
Welch column Xtimate C8 (3.0×250mm, 3μm) was used to analyze the VB12-related substance solution under this chromatographic condition. The chromatographic peaks were in the order of substance H, VB12, A, C, D, E and substance B, and the separation degrees were 30.3, 12.4, 2.2, 4.9, 1.6 and 4.0, respectively. The reproducibility verification results were good and met the analysis requirements.