Test Report
Chromatography conditions:
Column |
WM-CN100 (100m*0.25mm,0.2μm)07965-25011 |
|||||||||
Column Temperature |
|
|||||||||
Detector |
FID 250 ℃ |
|||||||||
Injection port |
250 ℃ |
|||||||||
Injection Volume |
1μL |
|||||||||
Flow Rate |
1.3 ml/min |
|||||||||
Carrier gas |
Nitrogen |
|||||||||
Split ratio |
30-1 |
|||||||||
Hydrogen gas |
32mL/min |
|||||||||
Air |
200mL/min |
|||||||||
Note |
/ |
Chromatogram and Data:
1) Mixture
Conclusion:
Using Welch WM-CN100 (100m*0.25mm, 0.2μm) 07965-25011 under this chromatographic condition, all peaks are normal, the peak order is consistent with the standard chromatogram, and the retention time is close to the reference, meeting the test requirements.