Holiday notice: We will be off for the Chinese Lunar New Year Holiday between Jan 25 and Feb 4. Responses may be delayed. Welch US and India operate normally.

Distinguishing Key Time Concepts in Chromatography Analysis

Distinguishing Key Time Concepts in Chromatography Analysis

Banner of this article

Chromatography is a widely used analytical technique that involves several key time-related concepts. These terms, including retention time, adjusted retention time, dead time, relative retention time, and delay time, play distinct roles in chromatographic analysis.

While all are expressed in terms of time, their implications differ significantly. Let’s explore these concepts to better understand their applications and distinctions.

Retention Time (RT, 𝑡𝑅​)

Retention time refers to the time elapsed from the injection of a sample to the appearance of the peak apex for a specific component on the chromatogram. Typically expressed in minutes (min), this metric is crucial for qualitative analysis and identifying target compounds.

For example, as shown in Figure 1 below, the component has a retention time of 10.0 min.

Figure 1. A chromatogram in which the compound with a retention time of 10.0 min.
Figure 1

Retention time can vary due to factors such as:

  • Buffer salt concentration and type in the mobile phase
  • pH and ratio of the mobile phase
  • Column temperature
  • Matrix effects

These small variations can significantly impact the separation of adjacent peaks, especially for sensitive analyses or complex samples.

In preparative chromatography, retention time is less significant. Overloading due to large sample injections often results in peak distortion (e.g., tailing or broad peaks), making it an unreliable metric in such scenarios.

Dead Time ( 𝑡0 )

Dead time represents the time required for the mobile phase (or unretained components) to pass through the column. It is the retention time of components that do not interact with the stationary phase. For instance, as shown in Figure 2, the dead time is 1.4 min.

Figure 2. A chromatogram identical to Figure 1, in which the dead time is 1.4 min,  thus the adjusted retention time is 8.6 min.
Figure 2

Adjusted Retention Time ( 𝑡′𝑅)

The adjusted retention time measures the extra time a component spends in the stationary phase due to adsorption or dissolution. It is calculated as:

t'R = tR - t0

In Figure 2, the adjusted retention time for the component is 8.6 min.

Relative Retention Time (RRT)

Relative retention time is not a time unit but a ratio used to compare the retention times of different components relative to a reference peak.

Chromatographic conditions often cause fluctuations in retention times. To prevent misidentification of peaks during analysis, a reference peak—typically a well-characterized or abundant component—is selected. The ratio of a component’s retention time to that of the reference peak remains consistent, even if retention times vary, enhancing reliability during peak identification.

Delay Time

A diagram illustrating "Delay Time" in high-pressure and low-pressure mixing systems

Delay time refers to the time required for the solvent mixture to travel from the mixing point (typically the mixer or proportioning valve in an HPLC system) to the column inlet. This time is influenced by the system’s dead volume, including the mixer and connecting tubing, and is also referred to as delay volume or gradient delay volume.

Delay time has a profound impact on accuracy and repeatability in liquid chromatography, particularly under gradient elution conditions. A larger delay volume prolongs the time it takes for gradient changes to reach the column, which can:

  • Delay peak elution
  • Cause peak broadening
  • Reduce sensitivity and column efficiency

Differences in delay volume across instruments can affect the reproducibility of gradient methods, potentially altering separation, retention time, and method transferability.

Conclusion

Understanding these time concepts is essential for interpreting chromatographic results accurately and ensuring robust method development. Proper awareness and control of these factors contribute significantly to achieving reliable and reproducible analytical outcomes.