Test Report
Chromatography conditions:
Column |
Welch Xtimate® UHPLC C18 2.1×150mm,1.8μm |
||||||||||||||||||
Mobile Phase |
|
||||||||||||||||||
Column Temperature |
40℃ |
||||||||||||||||||
Injection Volume |
5μL |
||||||||||||||||||
Flow Rate |
0.3 ml/min |
||||||||||||||||||
Note |
This test uses the matrix curve. |
Chromatogram and Data:
1) Blank
- 2) Sample solution
- (3) 10 ng/mL matrix curve
- (4) Sample spike, 10 ng/mL
Conclusion:
Using Welch Xtimate® UHPLC C18 2.1×150mm, 1.8μm, the recovery rates of beclomethasone, flunisolide, desonide, fluocinonide, flurandrenolide, prednisolone acetate, flumethasone, and fludrocortisone acetate in cosmetic water are 93.2%, 95.1%, 89.5%, 95.6%, 87.8%, 93.7%, 94.3%, and 102.3% respectively. The sample recovery rates meet the requirements.