Separation method development and reporting using Xtimate C18-DA

Separation method development and reporting using Xtimate C18-DA

Test Report


Chromatography conditions:

Column 

Xtimate C18(4.6×250mm,5μm)

Mobile Phase 

0.1% Formic Acid in Water: Acetonitrile

Injection Volume

5 μL

Flow Rate

1ml/min

Detector

DAD

Detect wavelength

254nm

Elution procedure

Time (min)

Mobile Phase A

Mobile Phase B

0

70

30

5

70

30

30

5

95

35

5

95

36

70

30

45

70

30


Note

N/A


  • 1. Spectra and data
  • (1) Full-range detection spectrum of sample solution


Signal: DAD1 A, Sig=254,4 Ref=360,100

Serial number

Retention time (min)

Peak Area

Area percentage

Resolution(USP)

Tailing factor

Number of plates(usp)

1

18.39

1014.28

30.80

 

0.81

200216.0

2

18.68

326.79

9.923

1.10

0.91

193280.3

3

20.04

229.68

6.974

7.83

1.07

205979.7

4

20.26

177.53

5.391

1.31

0.96

220968.7

5

21.93

378.27

11.49

9.02

0.76

198527.2

6

22.07

117.03

3.554

0.53

1.84

79144.9

7

22.47

1049.59

31.87

1.62

1.03

234886.7


(2) Local amplification detection spectrum of sample solution


Signal: DAD1 A. Sig=254.4 Ref=360,100

Serial number

Retention time (min)

Peak Area

Area percentage

Resolution(USP)

Tailing factor

Number of plates(usp)

1

18.39

1014.28

30.80

 

0.81

200216.0

2

18.68

326.79

9.923

1.70

0.91

193280.3

3

20.04

229.68

6.974

7.83

1.07

205979.7

4

20.26

177.53

5.391

1.31

0.96

220968.7

5

21.93

378.27

11.49

9.02

0.76

198527.2

6

22.07

117.03

3.554

0.53

1.84

79144.9

7

22.47

1049.59

31.87

1.62

1.03

234886.7


Conclusion: 

Using Welch Xtimate® C18 (4.6×250mm, 5μm) under this chromatographic condition, the four target peaks were separated from each other, meeting the customer's detection requirements