Separation of SK-SM1 and its isomers using Ultisil Amy-D

Test Report

Chromatography conditions:

Column

Ultisil® Amy-D(4.6×250mm,5μm)

Mobile Phase

n-hexane/isopropyl alcohol/trifluoroacetic acid=80/20/0.1

Detector

UV

Injection Volume

20 μL

Flow Rate

1.0 ml /min

Detection wavelength:

254 nm

Column Temperature

30℃

Note

/

Chromatogram and Data: 

  1. Detection amplification chart of mixed standard solution

Peak Name

Retention time(min) 

Area(mAU*s)

Height(mAU)

Resolution(EP)

Tailing Factor(EP)

Number of theoretical plates(EP)

SNR

SK-SM1

11.293

51106.878

1469.095

5.8

1.218

2291

12083.72

SK-SM1-Z

17.160

168.469

4.116

0.0

0.997

3906

33.859

Sum

 

 

 

 

 

 

 

 

2. Mixed standard solution detection full-scale spectrum

Conclusion:
Under these chromatographic conditions, Welch Ultisil® Amy-D4.6×250mm5μmcan meet the detection requirements.