How to Select a Suitable Column?

When you look up the pharmacopoeia method, you may find that it only stipulates the bonding phase of column and the information of other columns is not available. Therefore, you may feel confused about how to choose a suitable column. Welch Materials, as the most professional chromatographic consumables manufacturers in China, would like to share with you some experience as follows.

Basic Knowledge of Column

Chromatography is a separation analysis method, separation is the core, so the column which can achieve separation is like the heart of the chromatographic system. Common requirements for column are high column efficiency, good selectivity, fast analysis speed, etc. Packing materials for HPLC are mainly all-porous spherical silica gel, organic polymer matrix, core-shell silica gel, etc. The particle size is generally 3,5,7,10 µm, etc. And theoretical number of plates can reach 80,000 / m.

Selection of Size and Parameters of Packing Materials

Length of Column

Column Length Application
10~50mm Fast separation, 1.8 μm, 3 μm of packing materials are recommended
100~150mm Standard separation, 3~5 μm of packing materials are recommended
≥250mm High-resolution separation, 5~10 μm of packing materials are recommended

Inner Diameter of Column

Column Type Inner Diameter
Micro-column 0.3~1.0 mm
Short-diameter column 2.1 mm
Solvent-saving column 3.0 mm
Common analytical column 4.0, 4.6 mm
Semi-preparative column 7~10 mm
Preparative column >20 mm

Generally, 4.6 mm is the most commonly used for standard HPLC instruments. The 2.1 mm and 1.0 mm columns have higher sensitivity, but are more affected by the dead volume outside the column. Therefore, it is necessary to use a HPLC with low dead volume.

Particle Size of Packing Materials

Particle Size Performance
5 µm It can be used for separation from analysis to semi-preparation with widest application range
3–3.5 µm It is often used for rapid separation of analytical column, short column, saving solvent
< 2 µm It is often used for super fast separation with high resolution
5–10 µm It is often used to semi-preparative column
>10µm It is often used to preparative column

Column with small particle size of packing materials has higher column efficiency, sensitivity and shorter analysis time while column with large particle size of packing materials has lower column pressure and stronger anti-pollution capability.

Pore Size of Packing Materials

Pore size Performance
< 60 Å It is not useful for HPLC analysis, which will cause tailing and lower separation efficiency.
60 – 150 Å It is ideal for the separation of small molecules, such as drug molecules, traditional Chinese medicine and small molecule peptides.
300 –1,000 Å It is ideal for the separation of macromolecules, such as polypeptides, nucleosides and polymersl.
> 1,000 Å It is ideal for the separation of polymers, such as DNA and biological macromolecules.

The pore size of packing materials is mainly selected according to the molecular weight. Generally, compounds with molecular weight less than 2000 can match the packing material with a pore size of 100Å, while compounds with molecular weight greater than 2000 are recommended to match the packing materials with a large pore size of 300Å.

Carbon Load of Column

Ultisil Series C18 Column Carbon Load(%)
XB-C18 17
AQ-C18 12
LP-C18 10
PAH 22
Polar-RP 18
ALK-C18 12

1. The higher the carbon loading, the larger the k 'value, and the longer the sample retention time (the increase of mass transfer effect of stationary phase).

2. High carbon loading column is conducive to the separation of compounds that are not easy to be retained, meanwhile, it has good hydrolysis stability and reproducibility to improve the tailing of polar compounds.

3. Low carbon loading column is conducive to the analysis of neutral and alkaline compounds, which can reduce the loss of solvent and shorten the analysis time.

Classification of Packing Materials of Common Columns

Reversed-phase method

Ultisil® XB-C18 (Octadecyl or ODS)

Good universality, strong retention, wide application

Ultisil® XB-C8(Octyl)

Similar to C18, but with a slightly smaller retention value

UItisil® XB-C4(Butyl)

Has strong ability to retain hydrophobic and polar compounds

Ultisil® XB-C1(Trimethylsilane TMS)

Suitable for the separation of hydrophobic polypeptides and proteins

Ultisil® XB-PhenyI

Has good selectivity to aromatic compounds and polar compounds

Ultisil® XB-CN(Cyano)

Has unique selectivity to polar compounds

UItisil® XB-NH2(Amino)

The reversed mode is used to analyze monosaccharide samples.

Normal-phase method

UItisil® XB-CN(Cyano)

Good universality, moderate polarity, wide use

Ultisil® Diol

Its polarity is greater than CN and its stability and reproducibility are better.

UItisil® XB-NH2(Amino)

Analysis of polar compounds

UItisil® XB-SiO2

Suitable for the separation of polar compounds which are easily tailing.

Volume exclusion method

Xtimate® SEC molecular exclusion column (hydrophilic silicone globular protein)

Xtimate ® SEC molecular exclusion column (hydrophilic globular protein silica gel) Widely used in separation and determination of water soluble polymer and biological macromolecular.

Ion exchange method

Ultisil® XB-SCX(sulfonic acid group)

Strong cation exchange column

Ultisil® XB-SAX(quaternary ammonium salt) Strong anion exchange column

Xtimate ® Sugar-H(H+)

Widely used in the separation of organic acids and sugar

Xtimate ® Sugar-Ca(Ca2+)

Mainly used for determination of mannitol

Reversed-phase method

UItisil® Amy-D

Amylose tris coated chiral column

Ultisil® Cellu-D

Cellulose tris coated chiral column

HILIC hydrophilic chromatographic method

UItisil® HILIC-SiO2

Has long lifetime and strong stability for separating strong polar substances in circumstance of high aqueous phase

If you have any problem or require further information, please contact info@welchmat.com.